Table of Orders: separations, locations and lengths

m=order, blz=blaze wavelength, fsr=free spectral range,
d=order separation, l=length of free spectral range, dsp=dispersion

mblz (A)fsr (A)d (mm)d (asec)l (mm)dsp (A/mm)
1633505.421.50.40611.52610.961.96
1623527.021.80.40311.43411.031.97
1613548.922.00.39911.34411.101.99
1603571.122.30.39511.25611.172.00
1593593.622.60.39211.16711.242.01
1583616.322.90.38811.07811.312.02
1573639.423.20.38510.98911.382.04
1563662.723.50.38110.90211.452.05
1553686.323.80.37810.81711.532.06
1543710.224.10.37510.73111.602.08
1533734.524.40.37110.64611.682.09
1523759.124.70.36810.55911.752.10
1513784.025.10.36510.47411.832.12
1503809.225.40.36110.39311.912.13
1493834.825.70.35810.31011.992.15
1483860.726.10.35510.22612.072.16
1473886.926.40.35210.14212.152.18
1463913.526.80.34910.06312.242.19
1453940.527.20.3459.98212.322.21
1443967.927.60.3429.89812.412.22
1433995.727.90.3399.81812.492.24
1424023.828.30.3369.73912.582.25
1414052.328.70.3339.66012.672.27
1404081.329.20.3309.58112.762.28
1394110.629.60.3279.50512.852.30
1384140.430.00.3249.42612.952.32
1374170.630.40.3219.35113.042.33
1364201.330.90.3189.27313.142.35
1354232.431.40.3159.19513.232.37
1344264.031.80.3139.12113.332.39
1334296.132.30.3109.04613.432.40
1324328.632.80.3078.97013.542.42
1314361.733.30.3048.89713.642.44
1304395.233.80.3018.82513.742.46
1294429.334.30.2998.74913.852.48
1284463.934.90.2968.68013.962.50
1274499.035.40.2938.60714.072.52
1264534.736.00.2918.53514.182.54
1254571.036.60.2888.46514.292.56
1244607.937.20.2858.39314.412.58
1234645.337.80.2838.32514.532.60
1224683.438.40.2808.25614.642.62
1214722.139.00.2788.18714.772.64
1204761.539.70.2758.12014.892.66
1194801.540.30.2738.05115.012.69
1184842.241.00.2707.98315.142.71
1174883.641.70.2687.92015.272.73
1164925.742.50.2657.85415.402.76
1154968.543.20.2637.78715.542.78
1145012.144.00.2617.72415.672.81
1135056.444.70.2587.65915.812.83
1125101.645.50.2567.59615.952.86
1115147.546.40.2547.53516.102.88
1105194.347.20.2527.47416.242.91
1095242.048.10.2497.41316.392.93
1085290.549.00.2477.35016.542.96
1075340.049.90.2457.29216.702.99
1065390.450.90.2437.23416.863.02
1055441.751.80.2417.17617.023.05
1045494.052.80.2397.11817.183.07
1035547.453.90.2377.06417.353.10
1025601.754.90.2357.00717.523.14
1015657.256.00.2336.95117.693.17
1005713.857.10.2316.89817.873.20
995771.558.30.2296.84518.053.23
985830.459.50.2276.79218.233.26
975890.560.70.2256.74418.423.30
965951.962.00.2236.69418.613.33
956014.563.30.2216.64418.813.37
946078.564.70.2206.59619.013.40
936143.866.10.2186.54919.213.44
926210.667.50.2166.50619.423.48
916278.969.00.2156.46419.633.51
906348.670.50.2136.41919.853.55
896420.072.10.2126.37620.083.59
886492.973.80.2106.34020.303.63
876567.675.50.2096.30020.543.68
866643.977.30.2086.26220.783.72
856722.179.10.2066.22921.023.76
846802.181.00.2056.19421.273.81
836884.182.90.2046.16421.533.85
826968.085.00.2036.13721.793.90
817054.087.10.2026.10922.063.95
807142.289.30.2016.08022.334.00
797232.691.60.2006.05922.624.05
787325.493.90.1996.03922.914.10
777420.596.40.1996.02323.204.15
767518.198.90.1986.01223.514.21
757618.4101.60.1985.99423.824.26
747721.3104.30.1975.98424.144.32
737827.1107.20.1975.98124.484.38
727935.8110.20.1975.97924.824.44
718047.6113.30.1975.98025.164.50
708162.5116.60.1975.98525.524.57
698280.8120.00.1975.99625.894.63
688402.6123.60.1976.01226.274.70
678528.0127.30.1986.03326.674.77
668657.2131.20.1996.05827.074.85
658790.4135.20.2006.08627.494.92
648927.8139.50.2016.12527.925.00
639069.5144.00.2026.17028.365.08
629215.8148.60.2046.22028.825.16
619366.9153.60.2056.27929.295.24
609523.0158.70.2086.34629.785.33
599684.4164.10.2106.42430.285.42
589851.3169.90.2136.51230.805.51
5710024.1175.90.2166.60731.355.61