Kast Blue Dewar Specifications
Uses
Kast spectrograph, blue channel Date installed 9/20/2008 Detector
Fairchild CCD3041
2k x 2k
15 u-meter pixelsAmplifiers AR(UR),AL(UL),BR(LR),BL(LL) are
working. Temperature
-104 C Gain (e-/DN) 1.2 (slow)
1.3 (fast)Noise 3.7 e- (slow)
6.5 e- (fast)Dark current 0.02 e-/pixel/sec @ -50 C Horizonal CTE (-50 C) UL: 0.9999997
UR: 0.9999984
LL: 0.9999981
LR: 0.9999998Vertical CTE (-50 C) UL: 0.9999994
UR: 0.9999994
LL: 0.9999994
LR: 0.9999994Non-linearity (%) UL:0.04
UR: 0.02
LL: 0.01
LR: 0.00Full well > 109,000 e- (MPP)
Notes