| Uses |
Prime Focus Camera (3-m) |
| Date installed |
1/17/06 (replaces Site 2kx2k) |
| Detector |
Fairchild CCD486 back illuminated 4096x4096
15 u-meter pixels |
| Amplifiers |
AR(UR),AL(UL),BR(LR),BL(LL) are working. |
| Temperature |
-107 C. |
| Gain (e-/DN) |
1.08 slow 1.00 fast |
| Noise (e-) |
| sample time | gain setting | gain | noise |
| 8us (slow) | g1 | 1.08 | 4.1 e- |
| ?us (fast) | g? | ? | ? |
|
| Dark current | 7.4 e-/pixel/hour @ -75c (mpp)
5.5 e-/pixel/hour @ -120c (mpp)
3.2 e-/pixel/hour @ -132c (mpp)
|
| Serial CTE |
| UR: | 0.9999971 @ -65c
0.9999933 @ -90c
0.9999927 @ -100c
0.999991 @ -130c
|
|---|
| UL: | 0.99998 @ -130c |
|
| Linearity |
|
| Full well |
> 118 Ke- (mpp) |
| Fringing |
7.5% @ 8000A
11.3% @ 9000A
16.7% @ 10000A |