Detector Data for Mount Hamilton Facility Instruments

Dewar #2 Specifications

Uses 1-meter direct imaging
Date installed4/3/01
Detector LL3-w20-0,0 AR AL
thin 2K x 2K
15 u-meter pixels
reported by M. Wei
AmplifierAR
Temperature-125 C.
Gain (e-/DN) AR: 1.54
AL: 1.70
x-ray test
in Dewar #7
Readnoise AR: 11.8 e-
AL: 11.7 e-
Dark current (-125c) 0.5 e-/hr/pixel
Full Well MPP100,000 e-
non-MPP118,000 e-
Serial CTE AR: very good
AL: very good
Vertical CTEvery good
Notes Replaces unthinned, phosphor coated, 2k x 2k device.